Copper Pillar is rapidly being adopted as a bumped wafer interconnect. The construction is that of a Copper cylinder around 50µm in diameter and height, topped with a dome of solder.
As with any interconnect the quality of the bonding process between the different parts is vital for the reliability of the finished product. The bond between the copper and its pad on the wafer are of particular interest as this is seen as the most likely failure mode. In a bond test a failure mode between the copper and wafer is then the “Failure mode of interest”. If you are more interested in the solder to copper interconnect, click here.
This failure mode can be produced by either a pull or a shear test and the measured bond strength used for your process control. In many other applications pull testing is typically preferred because the bond is subjected to a simple tensile load, distributed over the bond area. The bond separation is clean making failure mode analysis of the surfaces relatively easy. Unlike solder bumps, Copper is relatively hard and gripping it therefore easier.
Solder balls require precise reforming in order to be able to apply a meaningful test load on to the bond. Copper also has to be reformed in order to be able to grip it but this takes the form of well know gripping methods such as plain surfaces and friction, a few serrations that slightly reform the copper in order to get a mechanical grip or a slight tapper.
An alternative to pull testing is a shear test. If this is done with a very low shear height the bond is subjected to mostly shear. If a high shear height is used the bond is loaded by shear and a bending moment. It is though the bending loads, and in particular the tension side of the bending moment, that typically cause the bond to fail. The pillar tends to pivot on its edge furthest from the shear tool, making the failure mode very similar to that of a pull test.
The advantages of a high shear test are that the tool and test are much simpler. Making pull tweezers with 50µm cavities is difficult and costly. Aligning them to the pillar takes time and requires skill. A shear test is no different to that commonly used when testing wire bonds. The only difference being the shear height is bigger. The tools cost a lot less and its alignment to the pillar is not so critical, making it less skilled and faster.
Variations in shear height will cause variations in the test result but with the Condor Sigma accuracy of ±1µm these affects are likely to be less than those caused by the grip reforming and operator misalignment in a pull test. It is though beneficial to test with the highest possible shear height as this reduces any effect from the small variations in height and assures that the bending moment dominates the failure mode.
For a solder ball a Cold Bump Pull test is known to be better than shear. A 1st bond Gold ball test is best done with a low shear height. The reason for both of these comes from our “Golden rules of bond testing” ©, they both produce the highest number of failure modes of interest or the highest possible test force.
In the case of Copper pillars we should be guided by these same rules. Shear is the simplest test and it does produce the failure mode of interest. The optimum shear height will be the one that maximises this or produces the highest test force.
If you are interested in the interconnect between the solder and the copper, it depends on the shape of the construction whether a Cold Bump Pull (CBP) type test is feasible. If not, a shear may be the only effective test to qualify your process.
More on CBP can be found in our how-to.
The test sequence for the CBP-approach is as follows:
As with any shear test, testing the solder to copper interconnect requires accurate alignment of the shear tool to the bond line. When the height of the bond line from the substrate varies due to tolerance build up conventional shear height methods can be a problem.
A solution is XYZTEC’s unique top landing shear method. The tool lands on the top of the solder, then moves back and down programmable amounts to guarantee a shear height relative to the top of the pillar rather than its base.
The video below shows a test. The programmed landing force is 5gf. As the landing contact is initially a point, even with a low landing force, a small amount deformation occurs on the top of the solder. The loads on the pillar bond though are small and compressive and have no detrimental effect on the measurement. The Condor Sigma is the only bond tester in the world with a programmable landing force and is capable of lower landing forces than any other tester.
The choice between pull or shear testing depends on the application and the test objectives. XYZTEC has a lot of experience with different types of copper pillar products and can advise you on how to test in order to get the best information for your quality assurance process. If you are interested to find out more please contact us for more information, to request an application report, a demonstration or a quotation.
The Condor Sigma is not only the most accurate bondtester in the world, but also the fastest. Click here to read the study that proves the Condor Sigma is up to 39% faster than the competition.
Please contact us if you have any questions or special bond testing requirements.